categories

JP-16001G METTLER SCALE

JP-16001G METTLER SCALE Common Process for Balance Limit values Maximum capacity 16200 g Readability 0.1 g Repeatability (at nominal load) 0.1 g Linearity deviation 0.2 g Sensitivity temperature drift 5 ppm/°C Typical values Repeatability (at nominal load) 0.07 g Minimum sample weight (acc. to USP) 210 g Minimum sample weight (U=1 %, k=2) 14 g Minimum sample weight OIML 5 g Settling time 2 s Internal adjustment yes, FACT Interfaces RS232C , USB Weighing technology MonoBloc Balance dimensions (W x D x H) 363x346x118 mm Weighing pan dimensions 351×245 mm Weight of balance 10.7 kg Weights for routin testing Large Weight/Class OIML/ASTM 10 kg / F2/4 Small Weight/Class OIML/ASTM 500 g F2/4 ..

JP-16001G METTLER SCALE

Common Process for Balance

Limit values

Maximum capacity

16200 g

Readability

0.1 g

Repeatability (at nominal load)

0.1 g

Linearity deviation

0.2 g

Sensitivity temperature drift

5 ppm/°C

Typical values

Repeatability (at nominal load)

0.07 g

Minimum sample weight (acc. to USP)

210 g

Minimum sample weight (U=1 %, k=2)

14 g

Minimum sample weight OIML

5 g

Settling time

2 s

Internal adjustment

yes, FACT

Interfaces

RS232C , USB

Weighing technology

MonoBloc

Balance dimensions (W x D x H)

363x346x118 mm

Weighing pan dimensions

351×245 mm

Weight of balance

10.7 kg

Weights for routin testing

Large Weight/Class OIML/ASTM

10 kg / F2/4

Small Weight/Class OIML/ASTM

500 g F2/4